Omega-engineering DAQ-12 Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Matériel Omega-engineering DAQ-12. Omega Engineering DAQ-12 User Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 31
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 0
INTERFACE CARDS FOR PERSONAL COMPUTERS
DAQ-12
Data Acquisition Adapter
For 16 bit ISA compatible machines
Users Manual
OMEGA ENGINEERING, INC. TEL: (203) 359-1660
One Omega Drive FAX: (203) 359-7700
P.O. Box 4047 Toll free: 1-800-826-6342
Stamford, CT 06907-4047 E-mail: [email protected]
http://www.dasieee.com
Vue de la page 0
1 2 3 4 5 6 ... 30 31

Résumé du contenu

Page 1 - Users Manual

INTERFACE CARDS FOR PERSONAL COMPUTERSDAQ-12Data Acquisition AdapterFor 16 bit ISA compatible machinesUsers ManualOMEGA ENGINEERING, INC. TEL: (2

Page 2 - WARRANTY/DISCLAIMER

2. Circuit Board Description and ConfigurationThe base address of the DAQ-12 is selected using switches SW1 and SW2. The operatingmode of the DAQ-1

Page 3 - Declaration of Conformity

The amplifier stage of the A/D converter circuit performs two functions: (1) amplifies lowlevel input signals and (2) converts this input signal into

Page 4 - http://www.omega.com

Table 2-1 details the available gain settings and resulting input ranges for the various input configurations. Note that the 'gai

Page 5

The final stage of the A/D converter circuit is the A/D converter IC. The converter must beconfigured for unipolar or bipolar input voltages as shown

Page 6 - 1. Introduction

n/a0000 1111 1111 1111100000 0111 1111 11110000 1000 0000 000050000 0100 0000 00000000 0100 0000 00002.50000 0000 0000 00000000 000

Page 7 - List of Figures and Tables

The D/A converter channels may also be operated in unipolar mode: 0 to +5 volts, or bipolarmode: -5 to +5 volts. The output mode is selected using jum

Page 8

2.3 Digital Input/OutputThe DAQ-12 offers four bits of digital output and four bits of digital input forcontrol/monitoring of external digital devices

Page 9 - 1.2 DAQ-12 Specifications

2.5 Clock SelectionThe DAQ-12 is equipped with a programmable clock circuit to produce data sampling ratesindependent from the clock rate of the host

Page 10

2.5.1 Internal ClockSampling rates for the internal clock can be calculated using the following equation: t = 100ns * [N1*N2] o

Page 11 - Enable Pre-Scaler

When configured for a 48-bit divider, the first sampling period will be slightly longer thanthe others because the first clock period is required to

Page 12

WARRANTY/DISCLAIMEROMEGA ENGINEERING, INC., warrants this unit to be free of defects in materials and workmanship for a period of 13months from the da

Page 13

When the multi-function timer is used as a pre-divider, the frequency of the external clockinput may be varied from DC to 10 MHz as long as the high p

Page 14

2.7 Direct Memory AccessDirect Memory Access (DMA) transfers provide a way of transferring data from theDAQ-12's A/D converter into the persona

Page 15

2.8 InterruptsThe DAQ-12 is capable of generating an interrupt from one of four sources:1. End of conversion signal2. DMA terminal count3. Multi-funct

Page 16 - 2.4 Base Address

3. External ConnectionsThe DAQ-12 is equipped with a high density 62-pin connector as shown in Figure 3-1.2 - CH1+ 4 - CH2- (CH10+)6 - CH4

Page 17 - 2.5 Clock Selection

CHx-, CHx+ : Analog inputs to the analog to digital converter. When using differential inputmode, eight input channels are available (CH0+ to C

Page 18

4. Register Description and ProgrammingThe DAQ-12 uses 16 consecutive I/O address locations in the range 0 to FFFFH. The cardutilizes these address

Page 19

INT2, INT1 and INT0 control the DAQ-12 interrupt source.INT2 INT1 INT0 DESCRIPTION0 0 0 Interrupt disabled1 0 0 Interrupt timer 21 0 1 Interrupt on

Page 20 - 2.6 Trigger Selection

VALID - when set, logic 1, indicates at least one data sample was lost because it was read bythe personal computer before the next sample was converte

Page 21 - 2.7 Direct Memory Access

4.1.4 D/A Converter 0 RegisterAn output to this register causes the lower twelve bits of data to be converted to an analogoutput on D/A converter chan

Page 22 - Factory default = IRQ 5

4.2 Programming the 8254 Counter/TimerThis section provides programming information for the 8254 counter/timer as implementedon the DAQ-12. For more

Page 23 - 3. External Connections

Declaration of ConformityManufacturer's Name: Omega Engineering, Inc.Manufacturer’s Address: One Omega DriveStamford, CT 06907-0047Application o

Page 24

Counter 1 - Clock rate register (high word)operating mode: 2minimum count value: 2configuration byte: 0 / 1 / 1 / 1 / 0 / 1 / 0 / 0 = 74H

Page 25 - 4.1 Register Description

DAQ-12 Users ManualVersion 2.20January 28, 1999Part No. 940-0030-220

Page 26

OMEGAnet On-line Service: Internet e-mail: http://www.omega.com [email protected] North Americ

Page 27

United Kingdom: One Omega Drive, River Bend Technology DriveISO 9002 CertifiedNorthbank, Irlam, ManchesterM44 5EX, England Tel: 44 (161) 777-6611FAX:

Page 28

Table of Contents294.2 Programming the 8254 Counter/Timer...284.1.9 8254 Control Word/Status Register...

Page 29

List of Figures and Tables25Table 4-1. DAQ-12 Address Map...15Table 2-3. D/A Converter Mode Select

Page 30

1. IntroductionThe DAQ-12 is a high speed data acquisition adapter for IBM AT and compatible machinesoffering eight differential analog input channel

Page 31 - Part No. 940-0030-220

1.2 DAQ-12 SpecificationsDACK 5, 6, 7DRQ 5, 6, 7DMA Levels:IRQ 2, 3, 4. 5, 6, 7, 10, 11, 12, 14, 15Interrupt Levels:0000H - FFFFHI/O Address Range:ISA

Commentaires sur ces manuels

Pas de commentaire