Omega-engineering DVG64 Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Ventilateurs Omega-engineering DVG64. Omega Engineering DVG64 User Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 2
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 0
DVG64
Electronic Vacuum Gauge
INSTRUCTION
SHEET
M3805/0502
DVG64 SPECIFICATIONS
Display Updated every second. 1 micron resolution below 200 micron of vacuum
Sensor Type Thermistor
Connector Type Standard 1/4 inch male flare fitting
Vacuum Range 0 - 12,000 Microns (0 - 1,600 Pascals) with vacuum increasing/decreasing indicator when above
12,000 Microns
Scale Microns, PSI, InHg, milliBars, Pascals, Torr, milliTorr
Operating Temp. Range 35°F to 125°F ( 2°C to 52°C)
Overpressure 300 PSI max (20 Bar)
Accuracy +/-10% (0 to 1000 microns)
Power Source 9 Volt Alkaline battery (not included)
Continuous Usage Over 35 hours
Auto Shut off After 10 minutes when vacuum reading is above 12,000 Microns (12 Torr)
Weight 6.7 ounces
Dimensions 5-1/2"H X 3"W X 1-1/4"D
FEATURES
• Reads vacuum in 7 international units
• Uses the battery power than other models
• Cleaning port provides quick cleaning and drying of sensor
• Uses standard 9V Alkaline battery
• Convenient built-in hanger
OPERATING CONTROLS
Turning the vacuum gauge ON: Press and hold the ON button
for about 3 seconds, until "POWER UP" appears on the display.
Turning the vacuum gauge OFF: Press the OFF button
To prolong battery life, when vacuum reading is above 12,000
Microns for about 10 minutes, the unit will automatically turn OFF.
Changing the scale: Press the Scale button to change the display
to the next scale. The scale order is: Microns, PSI, Inches of
mercury (InHg), milliBars, Pascals, Torr and milliTorr. DVG64 will
keep the scale settings even if the power is turned OFF.
UNDERSTANDING THE DISPLAY
• When the vacuum reading is above 12,000 Microns (1,600
Pascals), the first line of the display shows "Atm." The second
line displays a bar graph to indicate the direction in which the
vacuum is moving. When the bar graph is moving from left to
right, the pressure is increasing. When the bar graph is moving
from right to left, the pressure is decreasing. The speed of the
bar graph indicates how fast the pressure is increasing or
decreasing. The bar graph indicator may be inaccurate for a few
seconds after the evacuation of the system has begun.
The bar graph disappears if the vacuum does not change for
about 10 seconds. When the vacuum reading is below 12,000
Microns (1,600 Pascals), the vacuum in the selected units is
displayed.
CONNECTING TO THE VACUUM SYSTEM
The VG64 should only be connected to the vacuum system at
the vacuum port. The "Auxiliary Port" is primarily for cleaning
and should normally be closed with the supplied cap.
CLEANING THE VACUUM SENSOR
It is recommended that the sensor be cleaned periodically to
maintain unit accuracy . Oil and other contaminants reduce the
accuracy of the unit. Follow the instructions below for cleaning.
Close the Vacuum port with the supplied cap.
Open the Auxiliary port.
Use an eyedropper to pour about 2 teaspoons of ordinary
rubbing alcohol into the Auxiliary port.
Close the Auxiliary port with the supplied cap. Both the
Vacuum and Auxiliary ports should now be closed.
Shake the DVG64 unit for about 10 seconds. A slight
movement of the vacuum sensor in the case is normal and
does not affect the internal connection in any way.
Open both the Vacuum and the Auxiliary ports.
Empty the alcohol, and air dry the sensor.
Close both the Vacuum and the Auxiliary ports with the
supplied caps when the DVG64 is not used. This prevents
contamination of the sensor.
Vacuum (microns) Micron (mTorr) Step Pascal Step Bar Step InHg Step PSI Step
15000 - 8000 1000 100 1 0.05 0.02
8000 - 5000 500 50 0.5 0.02 0.01
5000 - 2000 250 25 0.25 0.01 0.005
2000 - 1000 50 5 0.05 0.002 0.001
1000 - 500 10 1 0.01 0.0005 0.0002
500 - 200 5 0.5 0.0005 0.0002 0.0001
200 - 0 1 0.1 0.0001 0.00005 0.00002
DVG64 STEP TABLE
Vue de la page 0
1 2

Résumé du contenu

Page 1 - Electronic Vacuum Gauge

DVG64Electronic Vacuum GaugeINSTRUCTIONSHEETM3805/0502DVG64 SPECIFICATIONSDisplay Updated every Ω second. 1 micron resolution below 200 micron of vacu

Page 2 - RETURN REQUESTS / INQUIRIES

It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations thatapply. OMEGA is constantly pursuing certification of its pro

Commentaires sur ces manuels

Pas de commentaire