Power and Measure Transmitter Mode
The CL531 supplies 15Volts or 24 Volts to the transmitter and displays the output in mA or % on the CL531 display. In read mode
press the EZ-Dial™ Knob twice to get into the feature options. Then press the EZ-Dial™ Knob to select mA PWR - M. Then turn
EZ-Dial™ Knob to select power range (15V or 24V). The EZ-Check switch and EZ-Dial knob allow rapid and fine control of loop
current.
Read Mode
The CL531 can read loop currents from 0-24 mA. The CL531 limits current in read mode to 25mA to protect the devices in
the loop from over voltage or over current conditions.
Transmitter Input
(Sensor)
(Process Signal)
(Simulated Input)
Receiver
Power Supply
0to60VDC
Typical
2-Wire
Transmitter
+IN- REF +OUT-
Transmitter Input
(Sensor)
(Process Signal)
(Simulated Input)
Receiver
Power Supply
0to60VDC
Typical
2-Wire
Trans mitter
+IN- REF +OUT-