OMEGA Model CYD201/CYD208 User’s Manual
Operation 3-5
3.6.2 Verifying SoftCal™ Operation
Check the status of SoftCal™ by holding
UNITS
for 2-3 seconds. If
SoftCal™ is not in operation, "-000-" displays.
If the first digit of the display is 1, SoftCal™ is set for below 28 K. If either
the second or third digit is 1, SoftCal™ is set for above 28 K.
3.6.3 Erasing SoftCal™ Compensations
When SoftCal™ compensations are erased, the thermometer returns to
normal operation.
1. Move the SET switch to position 1.
2. Hold
UNITS
until the front panel displays 0. This key is a toggle. If the
display moves in the wrong direction, release and press again.
3. Press
CAL ENABLE
on rear panel using a pen tip. The display changes
from 0 to "-SOF-". The alarm setpoint displays after erasing SoftCal™
compensation.
4. Move the SET switch to position 0.
Vue de la page 26